Technical Advance |





From the Departments of Molecular Cell Biology*
and Pathology,
Leiden University Medical
Center, Leiden, The Netherlands; and the Department of
Production,
Eurogentec S.A.,
Seraing, Belgium
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Until now more than 80 types of HPV have been identified. The types associated with diseases of the anogenital tract can be classified on the basis of phylogenetic relationship8 and of association frequencies with benign or malignant cervical lesions as high-risk types (HPV-16, -18, -31, -33, -35, -39, -45, -51, -52, -54, -56, -58, -59, and -66) and low-risk types (HPV-6, -11, -34, -40, -42, -43, -44). The most common HR types are HPV-16, -18, -31, -33, and -45.6,9 The most commonly found LR types are HPV-6 and -11.
Because the detection and type-specific classification of HPV infection by an in vitro viral culture test is not possible and serological tests are still ineffective, a molecular DNA diagnosis is needed. Direct hybridization-based assays, such as Southern blotting and in situ hybridization have been described, but these methods lack sensitivity and specificity. Signal amplification assays such as used in the hybrid capture assays10 for HPV are often applied. After recent improvements in this technique, it was shown that at the 1 pg/ml cut-off point (equivalent to 100.000 viral copies/ml) results agreed well with those of a polymerase chain reaction (PCR)-based assay.11
Because of the high-type specificity and sensitivity provided by target DNA amplification, the most widely used method for HPV detection in cervical cancers is based on PCR using either type-specific or general primers for the amplification.12-16 After the PCR amplification step, a number of methods can be used for classification based on sequence differences. These are based on type-specific oligonucleotide hybridizations such as for example the reverse line blot detection method, restriction fragment length polymorphism (RFLP), or, as the gold standard, sequencing.17-19 As a disadvantage in general, the above-mentioned methods are not suited for quantitative determination of viral load and require extensive processing after PCR.
The introduction of real-time reporting of PCR amplification with fluorogenic reporter probes such as TaqMan and molecular beacon probes provides quick and convenient tools for sequence variation detection (genotyping) and quantification.20,21 Here we designed molecular beacons recognizing common high-risk (HPV-16, -18, -31, -33, and -45) and low-risk (HPV-6, HPV-11) types and analyzed their performance in a general primed HPV-PCR in terms of specificity and sensitivity in a series of 53 formalin-fixed, paraffin-embedded cervical cancers that represented 21 different HPV types. Furthermore, we worked out a two-step HPV detection/genotyping strategy. In the first HPV step, samples are quantitatively analyzed for HPV-DNA using SybrGreen as a general HPV-amplicon reporter. In the second step, samples positive in the SybrGreen HPV-PCR are then analyzed for the presence of the seven prevalent HPV genotypes using the developed molecular beacon HPV-PCR. As internal reference, a ß-globin molecular beacon-PCR with identical PCR-cycling conditions as the HPV-PCR amplification is included in this strategy to assess DNA quality and to quantitate nuclear genome copy number of the samples.
This strategy, which is amenable to multiplexing, needs minimal amounts of tissue material and uses homogeneous assay formats that do not require processing after PCR, is well suited for quantitative detection of HPV-DNA, and for genotyping prevalent HPV types in cervical cancers.
| Materials and Methods |
|---|
|
|
|---|
Cell Lines
Cell lines containing different types of integrated HPV were obtained from the American Type Culture Collection (ATCC, Rockville, MD) or established by us. Extracted HPV-16 positive DNA from the SiHa cell line (ATCC no. HTB35) or from the CaSki cell line (ATCC no. CRL 1550), HPV-18 DNA from the HeLa cell line (ATCC no. CCL2), and HPV-45 DNA from the CC-10A cell line22 were used for dilution series and as positive controls. A B-lymphoblastoid cell line not containing HPV DNA was used as a negative control. DNA was isolated from cell lines according to a standard proteinase K-SDS procedure.22
Patient Samples
All patient samples were retrieved from formaldehyde-fixed paraffin-embedded tissue blocks collected between 1991 to 2000 and archived at the Department of Pathology, Leiden University Medical Center. From each paraffin block, four sections were cut. The first and the fourth sections were of 5-µm thickness, mounted onto glass slides, and hematoxylin and eosin stained for histological evaluation. The second and third sections of 10-µm thickness were collected in 1.5-ml Eppendorf tubes for DNA isolation as described.23 To prevent cross-contamination, microtome blades were thoroughly cleaned and mock-tissue blocks (paraffin only) were cut between blocks.
For evaluation of the specificity of the HPV molecular beacon-PCR, 57 marked samples, representing 4 mock tissue and 53 HPV-positive tissue blocks of patients with squamous cell (n = 34), adenocarcinoma (n = 5), and adenosquamous (n = 2) carcinoma of the uterine cervix and CIN lesions (n = 10) were used, as well as HPV-positive biopsies from skin (n = 1) and vulva (n = 1). The original HPV typing was performed on basis of general primer (GP) HPV-PCR followed by sequencing.23 The HPV type was not known to the investigators performing the molecular beacon PCR (KS, ESa). The four mock tissue blocks were included as negative controls.
The 53 patient tissue blocks represent 21 different HPV types (HPV-6, -7, -11, -16, -18, -26, -31, -33, -35, -45, -51, -52, -53, -55, -56, -58, -59, -66, -68, -70, and -73). Of the 53 patient samples, 32 contained HPV types matching with the seven molecular beacon probes used here (HPV-6, -11, -16, -18, -31, -33, and -45), whereas 21 samples represented the 14 other types. It should be noted here, that this set of 21 tissue samples is biased toward the infrequent types.
For evaluation of the false-positive rate of the general HPV SybrGreen-PCR assay, tissue blocks of 25 randomly selected ovarian cancers were used, which are considered to be negative for HPV.24
To assess the feasibility of a two-step HPV detection strategy (general primer HPV-PCR amplification reporting by SybrGreen, followed by molecular beacon HPV-PCR typing). Nineteen samples derived from 8 cases representing carcinoma of the uterine cervix (6 cases), 1 CIN II lesion, and 1 laryngeal papilloma were additionally selected. Cervical cancer cases were sampled from different paraffin blocks containing the tumor, histologically tumor-free lymph nodes, and tumor-free operation edges. Again, neither HPV types nor sample origin was known to the investigators who performed the SybrGreen- and molecular beacon-PCRs.
General Primer-Mediated HPV-PCR and Sequencing
General primer-mediated PCR (GP-PCR) and subsequent sequencing was performed for the detection and typing of HPV DNA. The ß-globin gene was used as an internal control for PCR amplification with the RS40/RS42 primer pair resulting in a 188-bp amplicon.25 All samples were subjected to GP-PCR with the HPV-consensus CPIIG/CPI primer pair to amplify a 188-bp fragment in the highly conserved E1 ORF region,13 the MY09/MY11 primer pair to amplify a 450-bp fragment in the L1 open reading frame,14 and the GP5/6 primer pair to amplify a 152-bp fragment in the L1 open reading frame.15 To determine the HPV subtype, PCR products were subjected to direct sequence analysis. Briefly, after purification of the PCR products using a Microspin G50-column (Amersham Pharmacia Biotech, Treiburg, Germany), the products were sequenced directly with the Big Dye Terminator kit (Applied Biosystems) using the PCR primers and a ABI Prism 377 automated sequencer. For nucleotide sequence analysis and comparisons, the programs Seqed, Fasta, and MAP of the Wisconsin Genetics Computer Group (version 9.1) sequence analysis software package were used. All available sequences of HPV types were retrieved from GenBank and complemented with sequences stored in a local HPV nucleotide sequence database. HPV types were identified when nucleotide comparisons revealed an identity of more than 95% with a known type.
Design and Characterization of Molecular Beacons
Molecular beacon probes were designed following the guidelines described by Tyagi and Kramer21 and in the related Internet-site (www.phri.nyu.edu/molecular_beacons). For a given HPV type the probe sequence was chosen to have maximum sequence difference with all other HPV types and with a melting temperature between 50 to 55°C. The intramolecular configuration of molecular beacons and the amplicon probing was modeled with the DNA mfold program that uses thermodynamic parameters established by SantaLucia.26 Sequences resulting in earlier opening of the stem because of an unexpected intramolecular configuration were excluded. Selected molecular beacon probes were further tested for absence of primer-probe, and probe-probe cross-hybridization.
Molecular beacon sequences (Table 1)
were
synthesized using standard phosphoramidite chemistry and purified by
high pressure liquid chromatography as described
before.27
Briefly, molecular beacons were synthesized
automatically with the 3'-DABSYL quencher linked to the solid phase
support (DABSYL-CPG, Glen Research). Oligonucleotide chain elongation
was run via standard cyanoethyl-phosphoramidite chemistry and
fluo-rophores (FAM, 6-carboxy-fluorescein; HEX,
hexacholo-6-carboxy-fluorescein; TET,
tetrahydrochloro-6-carboxy-fluorescein) were covalently linked to the
5'-end via iodoacetamide (Molecular Probes, Eugene, OR) or
phosphoramidite derivatives (Glen Research, Sterling, VA). After
release from the solid support, molecular beacons were purified twice
by reverse phase high pressure liquid chromatography using Waters 600E
equipped with a Waters 996 Photodiode array detector for simultaneous
detection at three different wavelengths. Ion molecular weights of
purified molecular beacons were determined by mass spectrometry using a
Dynamo Time-Off-Flight instrument.
|
Molecular Beacon-PCRs
PCR conditions for the ß globin gene PCR with the RS40/RS4225 and the HPV-PCR with the CPI/CPIIG primers13 amplification were identical. All PCR reactions were performed in the presence of 1x PCR buffer (15 mmol/L Tris/HCl, pH 8.0, 50 mmol/L KCl), 5 mmol/L MgCl2, 200 µmol/L of deoxynucleotide-triphosphate (dATP, dCTP, dGTP, dTTP), 1 U of Amplitaq Gold DNA polymerase (all from Applied Biosystems, The Netherlands), 250 mmol/L of each adequate primer, 200 nmol/L of each molecular beacon, and 50 µl of sample DNA in a 50-µl final volume. PCR amplification reactions were performed in an ABI Prism 7700 spectrofluorometric thermal cycler (Applied Biosystems, The Netherlands) with the following thermal cycle conditions: 95°C denaturation and enzyme activation for 10 minutes followed by 40 cycles of 95°C denaturation for 15 seconds, 50°C annealing for 50 seconds, and 72°C elongation for 1 minute. Fluorescence spectra were recorded during the annealing phase of each PCR cycle. The Sequence Detection Software (SDS v1.7) of the ABI-Prism 7700 was used to generate the amplification curves.
General HPV-PCR Reporting with SybrGreen
Amplification reactions were performed with 1x SybrGreen PCR Master Mix (Applied Biosystems), 200 nmol/L of CPII/G/CPI primers and 5 µl of sample DNA in a 50-µl final volume. Amplification reactions were performed in an ABI Prism 7700 spectrofluorometric thermal cycler (Applied Biosystems) with PCR conditions identical to the molecular beacon-PCRs. Fluorescence spectra were recorded during the elongation phase of each PCR cycle. The Sequence Detection Software (SDS v1.7) of the ABI-Prism 7700 was used to generate the amplification curves for each reaction.
To differentiate specific amplicons from nonspecific products (typically seen as a smear or primer-dimers on gel), a DNA association curve was generated after each reaction with the ABI Prism 7700 Sequence detection system (Applied Biosystems). After denaturation, the temperature was decreased from 95 to 25°C in 1°C steps with holds for 20 seconds. SybrGreen fluorescence was measured during each step and the first derivative of the fluorescence was plotted as a function of the temperature.
Dilution Series and Multiplexing
For generation of standard lines, dilutions were prepared by mixing SiHa and BSM cells such that between 6 and 6 x 105 copies of HPV-16 were present at a cell concentration of 1000 cells per 5 µl. Single, duplex, and triplex HPV molecular beacon PCR reactions were performed with SiHa, HeLa, CC10A cell lysates and HPV-16, -18, and -45 molecular beacons. All combinations were analyzed (three single, three duplex, one triplex). In addition, dilutions of the HPV-16 (5 x 103, 2.5 x 103, 103, 5 x 102 copies) with a constant background of HPV-18 (105 copies) and HPV-45 (1.2 x 105 copies) were prepared to assess the detection sensitivity and specificity in the presence of multiple HPV types.
Gel Analysis of PCR Reactions
For comparison and estimation of PCR efficacy, all samples amplified in this study were analyzed by standard agarose gel electrophoresis.
Calculations
The threshold cycle number (Ct) was calculated with Sequence
Detection System (SDS) software v1.7 (Applied Biosystems) and an
automatic setting of the baseline. The baseline value was the average
fluorescence value of PCR cycles 3 to 15 plus 10 times its SD. Standard
calibration curves were generated by plotting the Ct values against the
log of the input copy numbers. The Ct difference (
Ct =
CtHPV - Ctß-globin) was
taken as a relative measure of the virus DNA over nuclear genome copy
number.
| Results |
|---|
|
|
|---|
Our previous data showed that the CPI/IIG general primer PCR very
efficiently detected HPV-DNA in formalin-fixed paraffin-embedded tissue
derived from cervical cancers (Table 2)
.
This prompted us to design HPV molecular beacons for the 188-bp
CPI/IIG-bracketed HPV amplicon for five of the most prevalent HR
(HPV-16, -18, -31, -33, and -45) and two of the most prevalent LR HPV
types (HPV-6 and HPV-11) (Table 1)
. To determine any influence
of molecular beacons on the efficiency of the HPV-PCR, the
reactions were subjected to agarose gel electrophoresis and analyzed
for the presence of the 188-bp product. After breaking the seal at the
end of the study, the results were compared to our earlier findings
with HPV-PCR using CPI/IIG, MY09-11, GP56 primers (Table 2)
.
|
Molecular beacon probes fluoresce on hybrid formation with their
target sequences. The maximally discriminating hybridization
temperature is dependent on several factors such as GC content, probe
(loop), and stem length. To determine the optimal temperature range at
which each different HPV molecular beacon probe discriminates among the
HPV targets, melt-measurements of the various probe/target combinations
were performed. As an example, the melt characteristics of duplexes
between the HPV-16 molecular beacon and five target
HPV-oligodeoxynucleotide sequences are shown in Figure 1
. The smallest difference between the
HPV sequences used in this study is two nucleotides and exists between
HPV-16 and HPV-31 (Table 1)
. As shown in Figure 1
, at the PCR annealing
temperature of 50°C, the duplexes between the HPV-16 molecular beacon
and the HPV-31 target fluoresces only slightly more than the HPV-16
molecular beacon without template (dotted line). The other HPV-type
targets, because of their greater sequence variation, do not hybridize
at all over a large temperature range. The molecular beacons specific
for the other six HPV types were analyzed in a similar manner and none
of them showed hybrid formation with the nonmatching targets at a
temperature >50°C.
|
The four mock tissue samples showed no amplification pattern in the real-time HPV molecular beacon-PCRs. Also after gel electrophoresis no bands were visible.
The 21 tissue samples containing 14 different HPV types not matching with any of the selected HPV molecular beacons, did not report fluorescence amplification in any of the HPV molecular beacon-PCRs, but gave amplification of the 188-bp product as judged by gel electrophoresis (data not shown). Thus no aspecific reporting of molecular beacon probes was detected for the nonmatching types, demonstrating the high degree of specificity of molecular beacon probes.
Of the remaining 32 samples, 25 showed amplification reactions with one
and 2 showed amplification with two of the seven HPV molecular beacons.
The HPV types as determined by the molecular beacon HPV-PCR were in
accordance with the types obtained by the general primer PCR
amplification followed by sequencing. In Figure 2
, representative amplification curves
for the responding HPV molecular beacons are shown for three
HPV-positive patient samples (HPV-16, -18, and -45). All other
type-specific HPV molecular beacon-PCRs resulted in curves similar to
the one shown for the negative control and these never passed the
automatically set threshold.
|
Multiplexed HPV Molecular Beacon PCR
As a screening method, HPV molecular beacon PCR for seven HPV
types suffers from the drawback that an estimated 10 to 20% of total
HPV infection cases will not be reported.6,28
Expansion of
the number of types in the HPV molecular beacon-PCR would decrease this
false-negative rate. In view of the limited availability of sample DNA
and for assay convenience, multiplexing of HPV molecular beacon-PCR
then becomes desirable. To show the feasibility of multiplexing,
duplexed and triplexed HPV molecular beacon-PCRs were performed with
molecular beacons for HPV-16, -18, and -45 labeled with FAM, TET, and
HEX, respectively. Double (HPV-16 to -18, HPV-16 to -45, HPV-18 to -45)
and triple (HPV-16 to HPV-18 to HPV-45) infections were simulated by
mixing SiHa, HeLa, and CC10A DNA and treating such simulation samples
identical to the patient samples from the archive. As shown in Figure 3
, at approximately equal cellular DNA
input corresponding to
5 x 103,
105, 1.2 x 105
HPV-16, -18, and -45 molecules, respectively, an unequivocal detection
was observed in all cases.
|
General Prescreening Strategy by SybrGreen
Although the multiplexing capacity of molecular beacon PCR is higher than three,29 it is unlikely that it will approach the number of different HPV genotypes. For this reason we looked for a HPV-type-independent, homogeneous detection assay that could function as a general detection method for HPV prescreening. SybrGreen is a DNA intercalating fluorescent dye that quantitatively reports the presence of dsDNA. In combination with the CPI/IIG primer set, SybrGreen fluorescence can therefore be used in a homogeneous assay for quantitative monitoring of HPV-DNA amplification in real-time. SybrGreen will also detect any nonspecifically amplified DNA or primer-dimer artifacts. Nonspecific amplification products have a wide range of sizes, whereas a specifically amplified HPV product is of defined size, as are primer-dimer artifacts. These size differences of DNA fragments will result in differences in temperatures of maximum reassociation rate. Such information can be used to discriminate nonspecific amplification and primer-dimers from HPV-specific amplification.
In Figure 4A
the SybrGreen HPV-PCR
amplification patterns obtained with SiHa cells diluted in BSM cells
(corresponding to 50, 25, 12, and 2 copies of HPV-16 in a total of 1000
cells) are shown, whereas Figure 4B
shows the first derivatives of the
SybrGreen association curves. For the higher copy number inputs, clear
peaks can be discerned at 73°C. Only when the input HPV copy number
is as low as 2 per 1000 cells, a clear peak around 73°C cannot be
discerned anymore. Also in Figure 4
, the results obtained of the HPV
SybrGreen-PCR of the ovarian cancer cases are presented. In 22 out of
the 25 cases, no SybrGreen signal was observed after PCR cycle number
40, but three samples (12%) gave a SybrGreen signal over the threshold
at high-cycle number (n
36). As can be seen
in Figure 4B
, two of the three differentials of the SybrGreen
association curves show peaks quite distinct from the HPV amplicon,
whereas the third resembles the one of the lowest HPV input.
|
By gel electrophoresis, none of the ovarian cancer samples was positive. Hence, the false-positive rate of the HPV SybrGreen PCR assay is 12%. By combining it with the association analysis, the false-positive rate of SybrGreen analysis can be reduced to 4%.
Feasibility of the Two-Step HPV Detection and Genotyping Strategy
To assess the feasibility of the HPV detection and genotyping strategy, 19 additional samples, derived from eight patients were selected. ß-globin molecular beacon-PCRs, were performed and all samples showed a positive amplification pattern, confirming the suitability of the DNA for PCR amplification.
To determine the presence of HPV in samples, SybrGreen HPV-PCR was performed and 7 out of the 19 samples were positive. The seven positive samples also had peaks characteristic for HPV amplicons in the association analysis, including one histologically tumor-free lymph node (data not shown). Ten of the 12 samples negative in the SybrGreen analysis were from tumor-free operation edges or lymph nodes. The remaining two were cases of cervical cancer in which further analysis showed that these samples would not PCR amplify with the CPI/IIG primer set. Using other primer sets and sequencing, one of these cases was found to be HPV-18, whereas the other could not be typed.
In the HPV molecular beacon PCR, five of the seven SybrGreen-positive samples were positive: one (laryngeal papilloma) was genotyped as HPV-11 and four others (one adenocarcinoma and its tumor-free lymph node, two squamous cell carcinomas) as HPV-16. The three HPV SybrGreen-PCR-positive, but HPV molecular beacon-PCR-negative samples proved to be of a HPV type not included in the current HPV molecular beacon set (HPV-58 and HPV-68) based on data obtained by sequencing.
Quantitating Viral over Nuclear Genome Copy Number
Real-time monitoring of PCR amplification allows a reliable
determination of the input copy number using the Ct
value.30
Thus viral and nuclear genome copy numbers can be
determined, respectively, with the SybrGreen or HPV molecular beacon
PCR and the ß-globin molecular beacon PCR. As shown in Figure 2, a
good correlation was observed between the HPV SybrGreen and molecular
beacon PCR.
The difference in Ct values (
Ct) of the HPV and ß-globin PCRs
provides a measure of the viral DNA load per cell. Prerequisite for
this
Ct approach is that the efficiencies of the two PCR reactions
are approximately equal over a wide input range. As can be seen in Ct
versus SiHa cell input plots of Figure 5
, the lines for the ß-globin and HPV
SybrGreen-PCR run together throughout 4 decades. Thus these PCRs, which
run under identical cycling conditions, show equal efficiencies,
implying that the difference in Ct values (
Ct) may be used as a
measure of virus load per cell. From these data, it is obvious that the
present HPV detection strategy not only allows qualitative
determination, but also quantitation down to 50 cell equivalents with 1
to 2 copies of HPV per cell. For all cervical and ovarian samples
analyzed, Ct value for the ß-globin PCR ranged between 19.5 and 36.2.
For the cervical cancer cases, the Ct value of the HPV-PCRs ranged
between 18.2 and 35.8. The highest viral copy number was obtained with
a HPV-6-positive case with Ct values for HPV and ß-globin 18.2 and
31.5, respectively, indicating 104
HPV copy
number excess over the ß-globin gene. This finding was in line with
the histological findings that showed a small biopsy sample of a CIN
lesion.
|
| Discussion |
|---|
|
|
|---|
Molecular beacons were designed for typing the seven prevalent anogenital HPV types. On the basis of results obtained with 21 different HPV types represented in 53 formaldehyde-fixed samples, it can be concluded that the HPV molecular beacon probes are highly type-specific and well suited for application in a multiplexed homogeneous PCR assay for rapid genotyping of HPV sequences.
Because of the large sequence diversity of HPV types on one hand and the enhanced sequence variant discrimination of molecular beacons on the other,31 a single molecular beacon probe that detects all types is not feasible. To detect the presence of HPV in a type-independent manner and in a homogeneous format, we have evaluated the use of SybrGreen as a general reporter of HPV-PCR amplification. The false-positive rate of the HPV SybrGreen-PCR proved to be 12% (n = 25) when using signal-over-threshold as a criterion for positivity. However, the false-positive rate could be reduced to 4% by performing a PCR product-association analysis.
Techniques that use processing after PCR for detecting and genotyping
HPV are not only tedious, but also have limited capacity for
quantitating the (average) viral genome load per cell. Quantification
of DNA or RNA copy numbers, based on an external calibration curve
constructed from cloned or cell line DNA has a limited value for
samples with highly variable quality such as formaldehyde-fixed
tissues. This variability may cause 100 to 1000 times underestimation
of the actual copy number.32
The fact that the ß-globin
molecular beacon and HPV SybrGreen-PCR have identical efficiencies
(Figure 5)
enables one to use the
Ct value of these PCRs for an
HPV-type-independent quantitation of the average HPV copy number per
ß-globin gene that is also independent of the DNA quality. Tumor
samples are often contaminated by normal cells, the viral load per cell
determinations are therefore averages, implying that they must be
interpreted in the histological/cytological context of the sections
obtained before and after the cut for the DNA isolation.
The ß-globin molecular beacon-PCR for assessment of DNA quality and
nuclear genome quantity, the SybrGreen HPV-PCR for general HPV
detection and quantitation and the HPV molecular beacon-PCR for
genotyping the most prevalent HPV types allow an attractive
HPV-screening and -genotyping strategy, see Figure 6
. We showed the potential of this
strategy on eight new cases (19 samples). Intriguingly, one of the
samples was a histologically tumor-free lymph node that showed positive
amplification in the SybrGreen detection and that was subsequently
typed HPV-16 by the molecular beacon-PCR. This finding indicates the
sensitivity of our method, and shows its potential for detecting
minimal residual disease.33
|
Other studies describe detection of HPV in patient samples under homogenous (in casu closed tube) assay conditions. The strategy proposed here encompasses many of the favorable features of homogeneous detection assays described in these studies. Strauss and colleagues36 used general primer-based amplification reported by SybrGreen in real-time for a type-independent prescreening for the presence of HPV in samples, whereas Jordens and colleagues37 used molecular beacons coupled to a general primer, comparable to the Sunrise primer system38 to accomplish the same. In our strategy a similar prescreening is performed by SybrGreen reporting, whereas molecular beacons are used for the genotyping of the most prevalent high-risk and low-risk HPV types. It reveals the genotype information for the majority of the samples. Swan and colleagues39 used a hybridization probe-based assay with HPV-type-specific primer PCR and detection by a TaqMan probe in an end-point readout format. Although highly specific this approach has limited capacity for detection of multiple types and has also limiting quantitation potential. Joseffson and colleagues32 used the same TaqMan-probing principle in combination with a general primer pair and real-time read out, but did not incorporate general reporting for the remaining HPV types. The sensitivities of our method and the TaqMan approach are similar, but molecular beacon probes have a higher degree of specificity compared to TaqMan probes.40 Indeed, using the set of seven different probes on 21 types of HPV target we have not found any nonspecific reporting of nonmatching types nor in the included negative controls, contrary to the method described by Josefsson and colleagues32 in which the rate of false-positive reporting was found to be 10%.
The real-time PCR assay described here is a fast and simple method to quantitatively detect and type HPV DNA in cervical cancer. It can easily be extended to other HPV types and to tissue or cell samples.
| Acknowledgements |
|---|
| Footnotes |
|---|
Current address of E. S.: Department of Pathology, University Hospital Groningen, Groningen, The Netherlands.
Accepted for publication July 18, 2001.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
K. Biedermann, N. Dandachi, M. Trattner, G. Vogl, H. Doppelmayr, E. More, A. Staudach, O. Dietze, and C. Hauser-Kronberger Comparison of Real-Time PCR Signal-Amplified In Situ Hybridization and Conventional PCR for Detection and Quantification of Human Papillomavirus in Archival Cervical Cancer Tissue J. Clin. Microbiol., August 1, 2004; 42(8): 3758 - 3765. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Speich, C. Schmitt, R. Bollmann, and M. Bollmann Human papillomavirus (HPV) study of 2916 cytological samples by PCR and DNA sequencing: genotype spectrum of patients from the west German area J. Med. Microbiol., February 1, 2004; 53(2): 125 - 128. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. I. Brezinschek, N. Oppenheimer-Marks, and P. E. Lipsky Activated T Cells Acquire Endothelial Cell Surface Determinants During Transendothelial Migration J. Immunol., February 1, 1999; 162(3): 1677 - 1684. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |